ASTM F564-00
Historical Standard: ASTM F564-00 Standard Specification and Test Methods for Metallic Bone Staples
SUPERSEDED (see Active link, below)
ASTM F564
1. Scope
1.1 This specification covers characterization of the design and mechanical function of metallic staples used in the internal fixation of the muscular skeletal system. It is not the intention of this specification to describe or specify specific designs for metallic bone staples.
1.2 This specification includes the following four test methods for measuring mechanical properties of metallic bone staples:
1.2.1 Test Method for Constant Amplitude Bending Fatigue Tests of Metallic Bone StaplesAnnex A1.
1.2.2 Test Method for Pull-Out Fixation Strength of Metallic Bone StaplesAnnex A2.
1.2.3 Test Method for Soft Tissue Fixation Strength of Metallic Bone StaplesAnnex A3.
1.2.4 Test Method for Elastic Static Bending of Metallic Bone StaplesAnnex A4.
1.3 Unless otherwise indicated, the values stated in SI units are to be regarded as standard. The values given in parentheses are given for information only.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.
ASTM Standards
E4 Practices for Force Verification of Testing Machines
E122 Practice for Calculating Sample Size to Estimate, With Specified Precision, the Average for a Characteristic of a Lot or Process
E467 Practice for Verification of Constant Amplitude Dynamic Forces in an Axial Fatigue Testing System
F75 Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants (UNS R30075)
F86 Practice for Surface Preparation and Marking of Metallic Surgical Implants
F382 Specification and Test Method for Metallic Bone Plates
F565 Practice for Care and Handling of Orthopedic Implants and Instruments
F601 Practice for Fluorescent Penetrant Inspection of Metallic Surgical Implants
F629 Practice for Radiography of Cast Metallic Surgical Implants
Keywords
bending test; bone fixation; fatigue test; fixation devices; metallic bone staples; orthopaedic medical devices; pullout test; soft tissue fixation; surgical implants; Bone implant materials/applications--specifications; Constant amplitude bending fatigue test; Orthopaedic medical devices--specifications; Orthopaedic medical devices (bone)--specifications; Pull-out fixation strength; Soft tissue implant materials; Staples;
ICS Code
ICS Number Code 11.040.30 (Surgical instruments and materials)
DOI: 10.1520/F0564-00