Significance and Use
TEMPs may be composed of biological products (for example, human cells, organs, tissues, derivatives, and processed biologics), biomaterials (for example, substrates and scaffolds composed of
polymers or collagen), and biomolecules (for example, recombinant proteins, alginates, and hyaluronates) (see Terminology F2312 ). Those TEMPs that contain human viable cells, organs, or
tissues differ in terms of adventitious agent safety from other TEMPs because of the need to preserve viability of the organ, tissue, or cellular components. The need for preservation of
viability limits processing options for the reduction or elimination of adventitious agents. Examples of TEMPs are listed in Classification F2211 .
To ensure production and use of TEMPs with minimal risks associated with microorganisms and other adventitious agents, a multi-tiered approach is required. Donor testing, as well as testing
of components and raw materials by sufficiently sensitive assays that are state of the art is usually necessary. Compliance with good manufacturing practices (GMPs) and good tissue practices
(GTPs), where applicable, is required (21 CFR 210, 211, 820, 1270, and 1271). Although some of the components of the TEMPs may be processed to remove potential microbiological contaminants,
viable tissues and cellular components are generally unable to withstand rigorous processing without losing functionality. For those TEMPs containing tissues or cells for which banking is not
possible, even greater reliance on donor screening, component testing, and manufacturing controls is required. When more upfront testing is possible, there is generally greater confidence in
the safety of the final product. Process validation can enhance confidence in the ability of the TEMPs' producer to minimize risks from adventitious agents.
Throughout this guide, the reader is referred to other documents that may provide specific information that can be applied in the manufacture and testing of TEMPs. Although many of these
documents were not written with TEMPs in mind, parts are often applicable. Most of the potentially applicable position papers and guidance documents from many regions of the world can be
accessed via the internet. New documents are continually produced. The MDCA (U.K. Medical Devices Agency, now part of MHRA, Medicines and Healthcare Products Regulatory Agency) Code of
Practice for the Production of Human-Derived Therapeutic Products provides information on quality control, microbiological safety of donations, production, and processing practices. Two
Rijksinstituut voor Volksgezondheid en Milieu (RIVM) reports provide valuable information. One of these reports addresses preclinical safety assessment of TEMPs, and the other provides an
approach to risk management for TEMPs (1, 2) .
References may be made to draft guidances and rules. These should not be read as requirements.
1. Scope
1.1 This guide is intended as a resource for individuals and organizations involved in the production, delivery, and regulation of tissue engineered medical products (TEMPs). The safety from
contamination by potentially infectious adventitious agents is important in the development of all TEMPs as well as their components. This guide addresses how to assess safety risks
associated with adventitious agents and their byproducts. These agents currently include bacteria, fungi, mycoplasma, viruses, endotoxins, transmissible spongiform encephalopathies (TSEs),
and parasitic organisms. This guide does not address TEMPs with live animal cells, tissues or organs, or human cells, including stem cells, grown on any animal feeder cells. Also excluded is
patient follow-up testing.
1.2 This guide does not apply to any medical products of human origin regulated by the U.S. Food and Drug Administration under 21 CFR Parts 16 and 1270 and 21 CFR Parts 207, 807 and 1271.
This guide does apply to cellular therapies regulated under the PHS (Public Health Service) act.
1.3 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to
establish appropriate safety and health practices and to determine the applicability of regulatory limitations prior to use.
2. Referenced Documents (purchase separately) The documents listed below are referenced within the subject standard but are not provided as part of the standard.
Federal Regulations
21CFR1271 Human Cells, Tissues, and Cellular and Tissue-Based Products
MDA Standard
CodeofPractice for the Production of Human-Derived Therapeutic Products Available from Medicines and Healthcare Products Regulatory Agency (MHRA), Hannibal House, Elephant Castle, London
SE1 6TQ, U.K.
U. S. Pharmacopeia Document
UnitedStatesPharmaco (USP), Edition XXIV (24) Available from U.S. Pharmacopeia (USP), 12601 Twinbrook Pkwy., Rockville, MD 20852-1790, http://www.usp.org.
ASTM Standards
E1873 Guide for Detection of Nucleic Acid Sequences by the Polymerase Chain Reaction Technique
F2210 Guide for Processing Cells, Tissues, and Organs for Use in Tissue Engineered Medical Products
F2211 Classification for Tissue Engineered Medical Products (TEMPs)
F2312 Terminology Relating to Tissue Engineered Medical Products
F2386 Guide for Preservation of Tissue Engineered Medical Products (TEMPs)
ANSI/AAMI Standard
ST72 Bacterial EndotoxinTest Methodologies, Routine Monitoring and Alternatives to Batch Testing Available from American National Standards Institute (ANSI), 25 W. 43rd St., 4th Floor, New
York, NY 10036, http://www.ansi.org.
Keywords
adventitious agents; microorganisms; TEMPs; testing; Tissue engineered medical products (TEMPs); TEMPs (tissue engineered medical products); Adventitious agents;
ICS Code
ICS Number Code 11.100.99 (Other standards related to laboratory medicine)
DOI: 10.1520/F2383-11
ASTM International is a member of CrossRef.
ASTM F2383